Calibration Curve: Prepare Standard solution of Potassium (K) at 3, 6, 9, 12, 15 mg/L in 4% HNO3 and add Calcium (Ca) in all standard solution. |
|
|
|
Result: |
|
|
Figure 1 : Standard calibration for S-H BG correction (Absorption mode)
|
|
|
|
Figure 2 : Data of S-H BG correction (Absorption mode) |
|
|
Figure 3 : Standard calibration for D2 BG correction (Absorption)
|
|
|
|
Figure 4 : Data of D2 BG correction (Absorption mode)
|
|
|
Figure 5 : Standard calibration for Single beam mode (Emission)
|
|
|
|
Figure 6 : Data of Emission mode |
|
|
Summary |
The Self-reversal Background Correction (S-H) can suppress the matrix better than Deuterium Background correction (D2) because the value of BG (Abs) in the Self-reversal Background Correction is more than that of Deuterium Background correction The Self-reversal Background Correction is the good technique for determination the elements which the wavelength is more than 350 nm. |
|
The emission mode can measure the elements of Alkali metal such as K, Li, Na without Hollow Cathode Lamp A Calcium standard solution use to avoid an increase of the Potassium Chloride in the sample. |
|